Care
134
Operating Instructions Caleo, Software 2.n
Before Reusing With a Patient
Slide the bed out
z Open front door and fold it down.
1 Turn both knobs to the verticalposition marked .
2 Grasp bed by the recessed handle or by the knobs and pull
it out towards the front as far as it will go.
2 Push bed back until it clicks into place,
1 Turn both knobs to the position marked .
z Close front door.
If the bed cannot be pulled out or pushed in or if the knobs are
damaged:
z Take unit out of service.
To remove possible disinfectant residues, we recommend
running the incubator in standby mode.
z Switch on Caleo (see page 64).
z Activate air temperature control (see page 65).
z Run Caleo at 37
o
C with opened access ports.
If using a water container:
Do not refill the water container until just before placing a new
infant in the incubator (see page 49)!
If using a water bag:
Do not connect the water bag until just before placing a new
infant in the incubator (see page 50)!
WARNING !
Always ensure that the bed is pushed all the way in!
Otherwise the ducted flow of warm air will be interrupted,
and the infant may be warmed or cooled excessively.
018
1
1
2