Care
134
Operating Instructions Caleo, Software 2.n
Before Reusing With a Patient
Slide the bed out
z Open front door and fold it down.
1 Turn both knobs to the verticalposition marked .
2 Grasp bed by the recessed handle or by the knobs and pull
it out towards the front as far as it will go.
2 Push bed back until it clicks into place,
1 Turn both knobs to the position marked .
z Close front door.
If the bed cannot be pulled out or pushed in or if the knobs are
damaged:
z Take unit out of service.
To remove possible disinfectant residues, we recommend
running the incubator in standby mode.
z Switch on Caleo (see page 64).
z Activate air temperature control (see page 65).
z Run Caleo at 37
o
C with opened access ports.
If using a water container:
Do not refill the water container until just before placing a new
infant in the incubator (see page 49)!
If using a water bag:
Do not connect the water bag until just before placing a new
infant in the incubator (see page 50)!
WARNING !
Always ensure that the bed is pushed all the way in!
Otherwise the ducted flow of warm air will be interrupted,
and the infant may be warmed or cooled excessively.