Maintenance intervals
136
Operating Instructions Caleo, Software 2.n
Maintenance intervals
This chapter describes the required maintenance measures
to ensure correct functioning of the medical device. These
measures should be performed only by maintenance
personnel.
Maintenance terms and definitions
Safety checks
The safety-related checks do not substitute the maintenance
specified by the manufacturer, with the preventive replacement
of wearing parts.
1 Check the accompanying documentation:
– Current Instructions for use are available
2 Check all functions according to the Instructions for Use.
3 Check to ensure that the device combination is in flawless
condition:
– Labels are complete and legible
– No visible damages
CAUTION !
Disinfect and clean the incubator or the relevant parts before
each maintenance operation, even when returning the equip-
ment for repair purposes!
CAUTION !
Danger of an electric shock and device failure
Do not remove cover. Maintenance jobs must be performed
only by maintenance personnel. Dräger recommends the
DrägerService for this purpose.
Term Definition
Maintenance Combination of all technical and administrative measures during the life cycle of a medical device
for retention or restoration of working condition, so that the medical device can perform the required
function
Inspection Measures for the determination and assessment of the real condition of a medical device, including
determination of the causes of wear and tear, and derivation of the required consequences for
future use
Preventative
Maintenance
Recurrent specified measures intended to maintain the functional condition of a medical device
Repair Measures for the restoration of a medical device to the functional condition except for improvements
CAUTION !
Perform safety-related checks in the specified intervals. The
correct functioning of the medical device may be endangered
otherwise.