Maintenance intervals
137
Operating Instructions Caleo, Software 2.n
Safety checks
– Fuses accessible from outside match the specified
values
4 Check the equipment of the medical device fully in
accordance with the Instructions for Use.
5 Check electrical safety conforming to IEC 62353.
6 Check safety systems
Check the incubator temperature:
– Check the thermometer:
Measured value at the time of measurement should
not deviate more than ±0.8 °C from the displayed
atmospheric temperature
– Accuracy of thermometer < 0.1 °C
–
Steady state in the incubator:
Fluctuations of not more than ±0.5 °C over a period of
20 minutes
– Define measuring point:
10 cm above the center of the horizontal bed
– Carry out measurement:
at set value 36 °C, ambient temperature 23 °C ±2 °C