Maintenance intervals
138
Operating Instructions Caleo, Software 2.n
Maintenance
Maintenance
The following table shows the maintenance intervals:
CAUTION !
This device must be inspected and serviced at manufacturer-
specified intervals.
Interval times Who?
when
necessary
weekly every
2months
every
6months
once
a year
every
two years
Replaceable parts:
Fresh air filter X X User
Grommets X
1
1) Replace if the material becomes brittle or sticky or if strips of material have become detached
User
Fan motor X
2
2) Replace after 18000 – 20000 operating hours
Professionals
Lithium battery X Maintenance
personnel
O
2 sensors X Professionals
Skin temperature
sensors
X
3
3) At the latest when changing patient
User
Adhesive pads X
4
4) At the latest when changing patient
User
Mattress X User
Maintenance:
Inspection
X
Maintenance
personnel
Safety inspections
5
5) Designation is applicable in the Federal Republic of Germany; in the Republic of Austria this corresponds to the "Recurring safety-related
check"
X
6
6) Provided it is prescribed by national law, e. g., §6 MPBetreibV in Germany.
Maintenance
personnel
Measaurement checks
X
7
7) Provided it is prescribed by national law, e. g., §11 MPBetreibV in Germany.
Maintenance
personnel
Calibration:
O
2 sensors X Professionals
Scales X Professionals