Preparation
44
Operating Instructions Caleo, Software 2.n
Use of Doors, Ports, and Bed Adjusting Mechanism
To remove the canopy:
1 Grasp handles on the sides of the canopy with both hands.
2 Lift canopy horizontally off the supports.
Re-installing the canopy:
5 Place canopy horizontally, with the guide pins sliding into
the holes in the canopy supports.
Sliding out the bed
z Open front door and fold it down.
Pull the bed out:
3 Turn both knobs to the vertical position marked ,
4 Pull bed out towards the front as far as it will go, using the
recessed handle or the knobs.
z Upon completion of infant care procedures, push the bed
back in until it clicks into place. Turn knobs to the horizontal
position marked and close front door.
CAUTION !
Always take care not to damage sensor unit inside the
incubator when manipulating canopy !
WARNING !
Always observe the maximum load of the infant bed
(5 kg, 11 lbs).
WARNING !
Never leave infant unattended when the bed has been
pulled out to avoid any risk of an infant falling out of an
incubator.
WARNING !
Do not lean on the bed when it is pulled out. Equipment
damage with risk of patient injury may result.
WARNING !
Always ensure that the bed is pushed all the way in!
Otherwise the ducted flow of warm air will be interrupted,
and the infant may be warmed or cooled excessively.