Preparation
50
Operating Instructions Caleo, Software 2.n
Using Humidification Systems
Using a water bag
z Disinfect hands.
z Prepare a new transfer set (MX 17 018) and a water bag
with distilled water (e.g. Aqua dest.).
z Close the clamp on the transfer set.
z Insert the pin of the transfer set into the connector of the
water bag.
z Open the clamp on the transfer set.
z Bleed the transfer set (let sterilized water drain off).
z Close the clamp on the transfer set.
z Open the clamp on the transfer set.
z Switch on the humidification module on Caleo
®
and set the
humidity value (page 59).
Replacing the water bag:
Water bag empty = Water shortage alarm is triggered on
Caleo
®
.
z Disinfect hands.
z Close the clamp on the transfer set.
z Replace the water bag and reopen the clamp.
CAUTION !
Exclusively use sterilized distilled or demineralized water
for humidifying the incubator!
WARNING !
Use only authentic, originally sealed OEM bags
containing sterilized distilled or demineralized water.
Do not use any additives for water intended for
humidifying the incubator.
Do not use infusion solution bag for the humidification
system.