Preparation
51
Operating Instructions Caleo, Software 2.n
Integrated power strip
Integrated power strip
The integrated power outlet strip can be used to connect
infusion pumps and
–SpO
2 measuring equipment
as well as other equipment.
WARNING !
Output of the integrated power strip is not monitored!
Do not connect life support devices which do not have
their own power failure alarm.
Do not exceed the maximum permissible power input
for connected accessories
(all 4 sockets together: max. 2 A).
Do not exceed the maximum permissible total leakage
current. For the leakage current of Caleo without socket
strip see "Technical Data", page 166.
099