Checking Readiness For Operation
55
Operating Instructions Caleo, Software 2.n
Before Each Use
Before using the unit, make sure that the following tests
as well as the checks described under "Before Reusing
With a Patient", page 129, have been performed:
Disinfect hands before each test!
Check that access ports latch securely
Perform this test on all 4 access ports
1 Press down the ribbed area of the lever to open access port.
z Close access port until lever engages in the locked position.
z Try to open the access port by pulling it outwards by its edge
– it must not open.
If access port does not remain securely closed:
z Take unit out of service.
Check that large access doors latch securely
Perform this test on both doors
z Open door slightly.
2 Then, push door back into its closed position.
Turn the two knobs outwards until they engage in the
horizontal position.
If the front door fails to remain engaged or if the red marking is
visible or the double wall cannot be moved:
z Take unit out of service.
WARNING !
Always ensure that both knobs are engaged in position
in order to avoid any risk of an infant falling out of an
incubator.
The red catch behind each knob must no longer be
visible!
WARNING !
The attached double wall must be positioned parallel to
the front door otherwise the ducted flow of warm air will
be interrupted!