Checking Readiness For Operation
58
Operating Instructions Caleo, Software 2.n
Before Each Use
Start self test, check all displays and sound.
1 To switch unit on, press the on/off switch until it engages.
The incubator performs a self test.
It is the responsibility of the operator to verify screen displays,
LEDs, and audible alarms (enunciator and tone sequence).
The incubator will sound both an enunciator signal and
alarm tone sequence.
If one of the audible alarms does not sound:
z Take unit out of service.
Initially, screen and LEDs are dark, then illuminated.
If individual pixels on the screen or LEDs appear to be failing or
images are "burnt" into the screen:
z Take unit out of service.
The opening screen is displayed.
If no opening screen appears on the display:
z Take unit out of service.
The unit is now switched on.
WARNING!
The Caleo incubator is ready for operation only when all
checks have been performed successfully.
110
1
04237571