Operation
64
Operating Instructions Caleo, Software 2.n
Switching Caleo On
Switching Caleo On
z Connect unit to line power.
1 To switch unit on, press the on/off switch until it engages.
2 Position control unit so that its screen is clearly visible from
the working area.
An audible alarm is emitted.
– The opening screen is displayed.
The incubator performs a self-test.
It is the responsibility of the operator to verify screen displays,
LEDs, and audible alarm (see page 58).
– After the self-test, the standard screen for air temperature
control is displayed.
– The currently activated function is always highlighted by a
light background.
NOTE: In air temperature control mode, the unit takes
20 minutes to warm up. During this period, the alarm for
"Air temp. deviation above 1.5
o
C" is suppressed.