Operation
65
Operating Instructions Caleo, Software 2.n
Using Air Temperature Control
Using Air Temperature Control
Adjusting air temperature setpoint
1 To adjust setpoint, press key.
– The current (measured) value and setpoint appear on
screen both as bar graphs and numerical values.
– The message »set value with rotary knob« appears at the
top of the screen.
2 Turn rotary knob clockwise to increase setpoint.
2 Turn rotary knob counterclockwise to decrease setpoint.
2 Press rotary knob to confirm new setting.
If you do not wish to change the settings:
The display returns to the standard screen. The previous
setpoint is retained.
Or
– Wait for 7 seconds: Caleo emits 4 short beeps to prompt the
user to press the rotary knob. The display will immediately
return to the standard screen. The previous setpoint will be
retained.
WARNING !
Regularly measure infant temperature!
Do not leave canopy open for any length of time,
otherwise the air temperature inside the incubator will
drop.
Standard setpoint range 28
o
C to 37
o
C
Extended setpoint range 37.1
o
C to 39
o
C
20
o
C to 27.9
o
C
Default setpoint 33
o
C
Press and the new settings will be cancelled