Operation
65
Operating Instructions Caleo, Software 2.n
Using Air Temperature Control
Using Air Temperature Control
Adjusting air temperature setpoint
1 To adjust setpoint, press key.
The current (measured) value and setpoint appear on
screen both as bar graphs and numerical values.
The message »set value with rotary knob« appears at the
top of the screen.
2 Turn rotary knob clockwise to increase setpoint.
2 Turn rotary knob counterclockwise to decrease setpoint.
2 Press rotary knob to confirm new setting.
If you do not wish to change the settings:
The display returns to the standard screen. The previous
setpoint is retained.
Or
Wait for 7 seconds: Caleo emits 4 short beeps to prompt the
user to press the rotary knob. The display will immediately
return to the standard screen. The previous setpoint will be
retained.
WARNING !
Regularly measure infant temperature!
Do not leave canopy open for any length of time,
otherwise the air temperature inside the incubator will
drop.
Standard setpoint range 28
o
C to 37
o
C
Extended setpoint range 37.1
o
C to 39
o
C
20
o
C to 27.9
o
C
Default setpoint 33
o
C
Press and the new settings will be cancelled