Operating Instructions Caleo, Software 2.n
Using Air Temperature Control
Using the extended air temperature setpoint range
If the standard setpoint range is exceeded:
The advisory message
»confirm extended range with rotary knob«
will appear at the top of the screen.
z Press rotary knob to activate extended range.
z Turn rotary knob clockwise to increase setpoint further.
The advisory message » >37.0
o
C« appears on screen.
The following message appears at the top of the screen:
»set value with rotary knob«
z Press rotary knob to confirm the new setpoint.
The display returns to the standard screen. The measured
values are displayed.
The setpoint value and »set: >37.0 « are displayed
alternately.
The yellow bar LED lights up.
If the yellow bar LED starts flashing, the setpoint has been
increased by more than 1.5
o
C (see page 68), or other
alarms are active.
WARNING !
When the extended setpoint range for air temperature is
used, the infant’s temperature must be continuously
monitored.