Operation
76
Operating Instructions Caleo, Software 2.n
Using Skin Temperature Control
Adjusting the setpoint
1 Press key to adjust setpoint.
– The actual measured value and setpoint are displayed on
screen both as bar graphs and as numerical values.
– The following message is displayed in the top part of the
screen:
»set value with rotary knob«.
Standard setpoint range 34
o
C to 37
o
C
Extended setpoint range 37.1
o
C to 38
o
C
Default setting 36.5
o