Operation
78
Operating Instructions Caleo, Software 2.n
Using Skin Temperature Control
The display returns to the standard screen. The measured
values are displayed.
The setpoint and »set: > 37.0 « are displayed alternately.
The yellow bar LED lights up.
If the yellow bar LED starts flashing, the setpoint has been
increased by more than ±0,5
o
C* (see page 79), or other
alarms are active.
If you do not wish to activate the new setpoint:
1 Press button to cancel adjustment of the new setpoint.
The screen returns to the standard display mode and the
previous setpoint is retained.
Or
Wait for 7 seconds: Caleo emits 4 short beeps to prompt the
user to press the rotary knob. The display will immediately
return to the standard screen. The previous setpoint will be
retained.
* May be set within the range of 0.3 to 1.0
o
C. See page 108.