Operation
80
Operating Instructions Caleo, Software 2.n
Using Skin Temperature Control
If the sensor is defective:
– The screen displays the warning message
»Skin 1 sensor fault«.
– The intermittent alarm tone sequence (5 beeps) is sounded,
– The central alarm indicator lights up*,
1 3 dashes are displayed in place of a temperature reading.
2 The red bar LED starts blinking.
Then:
z Replace skin temperature sensor.
The alarm tone can be silenced for 5 minutes:
3 Press key to silence intermittent audible alarm,
or
4 press rotary knob.
– The warning message remains on the screen,
– The intermittent alarm tone is silenced,
– The central alarm indicator light goes out,
1 3 dashes continue flashing,
2 The red bar LED continues to flash.
For other alarms, see "Troubleshooting – Error Messages",
page 142.
See also "Alarm Description", page 161.
* The central alarm indicator light may be switched off.
See "Configuring system settings", page 106.
WARNING !
As long as 3 dashes remain on screen while the
incubator is operated in skin temperature control mode,
Caleo will not heat. Infant may become hypothermic.