Operation
82
Operating Instructions Caleo, Software 2.n
Using Humidity Control
Setting AUTO humidity
In AUTO mode, the humidity setpoint is automatically calcu-
lated and set by the system as a function of the air temperature
setting (see page 159).
The maximum relative humidity depends on the ambient
temperature and incubator air temperature at max. 75 % rela-
tive humidity.
1 Press soft key to switch humidity control to AUTO mode.
2 Press rotary knob to activate AUTO humidity.
The display returns to the standard screen. The measured
value and the automatic setpoint are displayed.
Manually adjusting the setpoint
3 Press soft key to switch humidity control to manual mode.
4 Turn rotary knob clockwise to increase setpoint.
4 Turn rotary knob counterclockwise to decrease setpoint.
4 Press rotary knob to confirm setpoint.
The maximum humidity reached depends on air temperature
and ambient humidity. At high air temperature or low ambient
humidity, the maximum humidity level that can be reached
inside Caleo will be reduced.
Standard setpoint range 30 % to 99 %
036
D
2
1
050
032
D
3
4