Loading...
Operation
92
Operating Instructions Caleo, Software 2.n
Selecting Menus
z Tubing grommets can be removed from the corner
segments of the hood so that hoses and cables connected
to the infant remain well organized during Kangaroo Mode.
Alarms
Alarm limits can be changed in the configuration
(see page 107).
If the skin temperature of the yellow skin temperature sensor
(skin 1) falls below the alarm limit set in the configuration:
The screen displays the warning message:
»Skin 1 temperature below 36.0
o
*
An alarm tone sequence (3 beeps) starts.
1 The central alarm indicator lights up**.
2 The measured value starts flashing.
3 The yellow bar LED starts blinking.
The intermittent audible alarm can be silenced for 15 minutes.
4 Press key to silence intermittent alarm tone
or
5 Press rotary knob.
The warning message remains on screen.
The intermittent alarm tone is silenced.
1 The central alarm indicator goes out.
2 The measured value continues to flash.
3 The yellow bar LED continues to blink.
When the measured value returns above the alarm limit:
The warning message disappears.
The intermittent alarm tone is cancelled.
1 The central alarm indicator goes out.
2 The measured value remains on screen without flashing.
3 The yellow bar LED goes out. If you are working in the
extended setpoint range, the yellow bar LED will stay
continuously lit.
For other alarms, see "Troubleshooting – Error Messages",
page 142.
See also "Alarm Description", page 161.
*The numerical values in this description are examples.
See "Configuring alarm settings", page 107.
** The central alarm indicator light may be switched off.
See "Configuring system settings", page 106.