Infinity M300/M300+ Features for VG2.4
14 Supplement – Infinity M300 and M300+ Series – VG2.4
To maintain a clear signal, change electrodes every 24 to 48 hours or more often when the following
occurs:
ECG signal degradation
Excessive patient perspiration
Skin irritation
Consider the following when selecting electrode sites:
Surgery keep electrodes as far from the surgical site as possible, while maintaining a clinically useful
lead configuration.
Burn patients use sterile electrodes. Clean the equipment thoroughly and follow hospital infection
control procedures.
ECG precautions
WARNING
Do not rely solely on the ECG when monitoring seizure-prone patients. Electrical artifacts of non-
cardiac origin, such as seizure, may prevent detection of certain arrhythmias.