Infinity M300+
Infinity M300+ specifications
Physical specifications
Size, H x W x D with battery
attached
157.5 mm x 81.1 mm x 30.6 mm (6.2 in x 3.19 in x 1.2 in)
Weight 310 g +/- 5g (10.93 oz)
Cooling Convection
Connections ECG, communication port for SpO2, programming cable
Environmental specifications
Ambient pressure Operating: 647 to 1060 hPa
Storage: 500 to 1060 hPa
Temperature Operating: 0 to 40 °C (32 to 104 °F)
Storage: –20 to +60 °C (–4 to +140 °F)
Humidity (non-condensing) Operating: 10 % to 85 %
Storage: 10 % to 85 %
Protection against ingress of
water
IPX ratings:
– Main unit with battery attached is IPX 4, protected against
splashing water.
– Main unit without battery attached is IPX 7, protected against
temporary immersion.
– Swappable battery alone, unattached, is not IPX rated.
Electrical characteristics
Power source Rechargeable 3.6 V lithium-ion swappable battery, available
from Dräger
Protection class Internally powered
Mode of operation Continuous
Estimated battery operating
time
ECG only: 22 h
ECG and continuous SpO2: 18 h
Duration of operation varies according to use of display, alarms,
and wireless environment (roaming).
NOTE
The duration of the battery charge for the M300+ may be
reduced by characteristics of the wireless environment such as
signal quality, power levels, power-save mode, etc. The battery
operating life can also gradually diminish over time.