Supplement – Infinity M300 and M300+ Series – VG2.4 17
Infinity M300/M300+ Features for VG2.4
Alarm settings
While the battery is attached to the M300+, alarm settings and current patient data remain intact for
approximately 80 hours after the unit powers down due to a depleted battery.
Low battery alarm
When the battery power level has reached approximately 30 minutes of remaining charge, a Low battery
message is reported visually and acoustically at both the ICS and M300/M300+ to indicate that the battery
is low. The M300/M300+ displays the Low battery status once every three minutes, plus or minus five
seconds.
Critical battery alarm
Displays a critical battery alarm message, Recharge battery, to indicate a low battery when the
M300/M300+ power level has reached approximately 15 minutes of remaining charge. The alarm is
reported visually and acoustically at both the M300/M300+ and the ICS. After the user acknowledges the
alarm, the message will rotate through any other displayed M300/M300+ messages.