Supplement – Infinity M300 and M300+ Series – VG2.4 19
Infinity M300+
M300+ swappable battery
The new Infinity M300+ operates with a swappable, externally rechargeable battery with the following
specifications:
Battery recharging time of approximately 3.5 hours
Battery swaps retain clinical data and current alarm settings for approximately two minutes
Battery swaps do not require tools
Battery provides a four LED charge level indicator when the battery is detached
The complete unit, with the swappable battery attached to the monitor, is rated IPX4, protected
against splashing water
Battery safety information
Observe the following warnings and cautions for the swappable battery.
WARNING
Do not try to power up the M300+ without a battery installed. Attempting to power up the M300+
without a battery installed will result in erasing any stored patient information from the M300+.
NOTE
IPX ratings:
Main unit with battery attached is IPX 4, protected against splashing water.
Main unit without battery attached is IPX 7, protected against temporary immersion.
Swappable battery alone, unattached, is not IPX rated.
WARNING
During battery replacement, there is no data monitoring and no transmission of data to the ICS.
However, clinical data and current alarm settings are retained in the M300+ for a maximum of two
minutes. Exceeding this time limit will result in loss of clinical data and alarm settings.
WARNING
Risk if the swappable battery is not replaced properly.
If the swappable battery is not reattached or serviced and replaced properly, short circuits and
high temperatures leading to explosion or fire may occur.