Supplement – Infinity M300 and M300+ Series – VG2.4 21
Infinity M300+
Preventive Maintenance
The following table shows the preventive maintenance interval for the M300+ swappable battery:
Caring for the M300+ lithium-ion swappable battery
The following table provides care instructions for the M300+ swappable battery.
Charge status for swappable battery attached to the M300+
The M300+ LED indicates battery charge status as follows:
Component Interval Measure Personnel Responsible
M300+ swappable
battery
Every 12 months Replace swappable
battery
Users (Medical or Technical
Personnel)
Activity When to Perform
Visually inspect the swappable battery Prior to inserting the swappable battery into the M300+
Clean the swappable battery – When the swappable battery is exposed to contami-
nants
– After each patient discharge
Refer to "Manual disinfectant and simultaneous cleaning"
on page 43 for more information.
Charge stored swappable battery When not in use for an extended time
Replace the swappable battery Once every year.
LED Battery charge status
Blinks green every 5 seconds Power on
Blinks amber every 5 seconds Low battery charge
WARNING
Do not immerse the battery in liquids as this
could cause a hazard.
NOTE
The duration of the swappable battery charge for
the M300+ may be reduced by characteristics of
the wireless environment such as signal quality,
power levels, power-save mode, etc. The battery
operating life can also gradually diminish over
time.