Supplement – Infinity M300 and M300+ Series – VG2.4 31
Infinity M300+
Accuracy SpO2: 0 to 69 % not specified
70 to 100 % sensor-specific as follows:
Masimo LNCS Sensors
SpO2 accuracy
1, 2, 3, 4
:
LNCS DCI, LNCS DCIP, LNCS Adtx,
LNCS Pdtx Not specified
Pulse rate accuracy: ±3 bpm or ±3 %,
whichever is greater
Low perfusion accuracy, SpO2: ±2 %
Low perfusion accuracy, pulse rate: ±3 bpm or ±3 %,
whichever is greater
Masimo RD SET Sensors
SpO2 accuracy
1, 2, 3, 4
:
RD SET Adt, RD SET Pdt, ±2 %
Pulse rate accuracy: ±3 bpm or ±3 %,
whichever is greater
Low perfusion accuracy, SpO2: ±2 %
Low perfusion accuracy, pulse rate: ±3 bpm or ±3 %,
whichever is greater
Nellcor Sensors
SpO2 accuracy
1, 2, 3, 4
:
OxiMAX MAX-A, OxiMAX MAX-AL,
OxiMAX MAX-P, DS100A ±3 %
Dräger Sensors
SpO2 accuracy
1, 2, 3, 4
:
MS16444 disposable foam pedi,
MS16445 disposable foam ad,
MS16449 disposable vinyl ad,
MS16448 disposable vinyl pedi,
MS13235 reusable sensor ±2 %
Infinity M300+ specifications