Reprocessing of device-specific components
44 Supplement – Infinity M300 and M300+ Series – VG2.4
Dräger recommends using a surface disinfectant from the list above. Other disinfectants are used at one’s
own risk.
Dräger points out that oxygen-releasing agents and chlorine-releasing agents may cause color change in
some materials. Color change does not indicate that the product is not functioning correctly.
At the time of validation, the listed surface disinfectants showed good material compatibility. The manu-
facturers of the surface disinfectants have verified at least the following spectra of activity:
– Bactericidal
– Yeasticidal
– Virucidal or virucidal against enveloped viruses
Observe the specifications of the surface disinfectant manufacturers. Use surface disinfectants that are
nationally authorized.
Surface disinfection with cleaning
1 Remove soiling immediately. Use a cloth dampened with disinfectant to remove soiling.
2 Perform surface disinfection.
3 After the product has been exposed to the disinfectant for the specified contact time, remove residual
disinfectant.
4 Wipe with a cloth dampened with water (preferably drinking-water quality). Allow the product to dry.
5 Check the product for visible soiling. Repeat steps 1 to 5 if necessary.
6 Check the product for visible damage and replace if necessary.
Quaternary ammonium com-
pounds
Mikrozid sensitive liquid
1)
Schülke & Mayr
Mikrozid sensitive wipes
1)
Mikrozid alcohol free liquid
1)
Mikrozid alcohol free wipes
1)
acryl-des
1)
Aldehydes Buraton
10 F Schülke & Mayr
1) Virucidal against enveloped viruses
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Ensure that no liquid penetrates the device.
Class of active ingredient Surface disinfectant Manufacturer