EasyManua.ls Logo

Dräger LFB PSS 7000 - Page 50

Dräger LFB PSS 7000
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
476 Issue date: 17 November 2006 50 of 80
20.60 Wait 30 seconds pre-alarm sounds do not cancel wait 15 seconds - full alarm sounds
Check LEDs, two blue, two red and one green illuminate alternatively Insert Bodyguard key
to cancel.
20.61 Test DSU by pressing the yellow button in the centre of the unit to
start the manual DSU alarm check LEDs, two blue and two red
illuminate alternatively. Remove and reinsert Bodyguard key to cancel.
’B’ test: leak test
20.62 To undertake the leak test the wearer will need to access the leak test
icon.
20.63 Press the left button of the Bodyguard unit. The unit will emit a single audible
bleep’, and then the ‘self-check’ sequence will start as described in (see image right).
20.64 The initial short display quickly changes to the LCD check display, the backlight
will momentarily illuminate and the blue, red, and single green LED’s in the lower
display cluster will momentarily illuminate (see image right).
Note: during each display the outer radial segments will initially illuminate and then
begin to switch off clockwise. If no action is taken by the time the last segment
disappears, the low pressure icon is momentarily displayed followed by the ‘self-
check’ icon which will restart.
20.65 When the leak test icon appears press the left button (see image right).
20.66 ‘Open cylinder’ icon momentarily appears (see image right).
20.67 The Bodyguard unit will display the ‘close cylinder’ icon and then
the ‘press right button’ icon; these two icons will alternate (see
image right).
20.68 Close the cylinder valve and then press the right button to confirm.
’B’ test: stabilisation
20.69 Upon the closing of the cylinder valve the Bodyguard unit will take a few seconds
to stabilise the air trapped between the cylinder valve and the first breath button on
the LDV (see image right).
20.70 The ‘leak test timing’ icon, is displayed throughout the leak test period of
approximately one minute. During this time the bar segments will switch off
clockwise (see image right).
Note: during the leak test there may be a drop in pressure displayed. A pressure
drop of up to 10 bar is acceptable and may be caused by the piston in the reducer
settling.
20.71 On completion of the leak test one of the following icons will be displayed:
(a) Leak test pass. This flashing icon indicates that the leak test was

Related product manuals