13 Accessories
Accessories are not part of PFG 07 G 003.
CO
2
Cl
2
H
2
S LC OV
3)
OV-A
3)
Indication range
0 ... 5 vol.-% 0 ... 20 ppm 0 ... 100 ppm 0 ... 200 ppm 0 ... 200 ppm
Calibration concentration 2.5 vol.-% in air 5 ppm in N
2
20 ppm in N
2
20 ppm in N
2
20 ppm in N
2
Temperature range, operation –20 ... 40
o
C
–4 ... 104
o
F
–30 ... 50
o
C
–22 ... 122
o
F
–40 ... 50
o
C
–40 ... 122
o
F
–20 ... 50
o
C
–4 ... 122
o
F
–20 ... 50
o
C
–4 ... 122
o
F
Alarm Threshold A1
2)
0.5 vol.-% 0.5 ppm 5 ppm 10 ppm 10 ppm
acknowledgeable Yes Yes Yes Yes Yes
latching No No No No No
Alarm Threshold A2
2)
1 vol.-% 1 ppm 10 ppm 20 ppm 20 ppm
acknowledgeable No No No No No
latching Yes Yes Yes Yes Yes
TWA Threshold A1
2)
0.5 vol.-% 0.5 ppm 5 ppm No No
STEL Threshold A2
2)
2 vol.-% 0.5 ppm 5 ppm No No
No. of STEL periods 4 4 4 No No
Average STEL duration 15 minutes 15 minutes 15 minutes No No
Warm up time 12 hours 30 minutes 5 minutes 18 hours 18 hours
Reproducibility
Zero point: ± 0.3 vol.-% ± 0.05 ppm ± 0.1 ppm ± 3 ppm ± 5 ppm
Sensitivity: [% of measured value] ± 20 ± 2 ± 5 ± 5 ± 20
Drift (20
o
C)
Zero point: ± 0.2 vol.-%/a ± 0.2 ppm/a ± 0.2 ppm/a ± 5 ppm/a ± 5 ppm/a
Sensitivity:
[% of meas. value/month
] ± 15 ± 2 ± 1 ± 2 ± 3
Sensor Order Number
1)
68 10 889 68 10 890 68 11 525 68 11 530 68 11 535
Sensor Data Sheet Order Number 90 23 923 90 23 924 90 23 970 90 23 994 90 23 995
Please be aware of possible sensor cross sensitivities (see DrägerSensor and gas detector manual at www.draeger.com/
sensorhandbook).
1) The sensors have a limited life time. Excessive storage reduces operating time of sensors.
The adequate temperature range for storage is 0 to 35 °C (32 to 95 °F).
2) Please be aware of special settings by customer requirements.
3) For ethylene oxide only.
Description
Order-code
Communication module, complete with USB cable
83 18 587
Calibration Adapter 83 18 588
Lithium battery 45 43 808
Dust and water filter 45 43 836
Leather carrying case 45 43 822
Bump Test Station, complete with test gas cylinder 58 L (gas type by customer request) 83 18 586
Dräger X-dock 5300 Pac
83 21 881