26 Dräger PEX 1000
Operation
Cable length between PEX 1000 and controller unit:
Core cross-section
Minimum
supply voltage
0.5 mm
2
(36 Ohm/km)
0.75 mm
2
(24 Ohm/km)
1.0 mm
2
(18 Ohm/km)
1.5 mm
2
(12 Ohm/km)
18 V 200 m 310 m 419 m 620 m
20 V 270 m 410 m 550 m 830 m
22 V 340 m 520 m 690 m 1040 m
24 V 410 m 620 m 830 m 1250 m
26 V 480 m 730 m 970 m 1469 m
28 V 550 m 830 m 1110 m 1679 m