5.3 Cleaning, Disinfecting and Drying
5.3.1 Important Information
Dräger recommend that contaminated and dirty components or
assemblies are carefully cleaned and disinfected then thoroughly dried
after use, or as and when considered necessary by the user.
To ensure correct operational condition of the equipment use the
cleaning and disinfecting solutions recommended by Dräger. The use
of any other product will invalidate the Dräger warranty and guarantee.
When using baths to contain cleaning and disinfecting solutions the
immersed components and assemblies must only be agitated manually.
Any form of mechanical agitation, e.g. ultrasonic, washing machines,
dishwashers e.t.c. must be approved and covered by a written
agreement with Dräger.
Note:
Cleaning solutions will eventually remove lubricants from moving parts.
Following any cleaning, disinfecting and drying carry out functional
tests. Re-lubricate moving parts as required.
Safety Note: When using cleaning and disinfecting agents refer to the relevant
manufacturers’ usage instructions. It is important that attention be paid
to concentration and reaction times. Do Not use organic solvents such
as Acetone, Alcohol, White Spirit, Trichloroethylene or similar.
Dräger recommends the following:
1. Cleaning:
Safety Wash
1 Litre Dispenser 3380164
1 Litre Refill 3380165
5 Litre Dispenser 3380166
5 Litre Refill 3380167
Sekusept
4 bottels @ 2 litres 7904071
These products must be used with cold water however Never exceed a
temperature of 30 degrees Celsius.
Note:
Rinse off cleaning solution in clean water before disinfecting.
2. Disinfecting:
Wipex Cloths (for manual disinfecting)
Pack of 50 3380375
Incidur (6 litre container) 7904072
Incidur (30 litre container) 7904073
This product must be used with cold water however Never exceed a
temperature of 30 degrees Celsius.
Note:
Rinse off disinfecting solution in clean water before drying.