Operation in Measurement Mode
27
Operating Instructions Vamos, 1. US ed.
Selecting Other SpO2 Averaging Intervals
Setting the Volume of the Pulse Tone/Alarm Tone
Selecting Other SpO2 Averaging Intervals
The user can choose between the following three averaging
intervals:
normal, slow, fast.
Upon switching Vamos on, the default setting is "normal".
Depending on the degree of agitation of the patient, two other
averaging intervals can alternatively be selected:
Long averaging interval: "slow"
The long averaging intervals are suitable for patients with
increased movement activity.
NOTE: With the "slow" averaging interval, no pulse rate is
displayed (Pulse ––) and the pulse tone is muted.
Short averaging interval: "fast"
The short averaging intervals are suitable for situations in
which the patient is calm.
To set:
Turn or press dial knob to select main menu.
In the main menu, turn dial knob to select »SpO2«.
Press dial knob to confirm.
Turn dial knob top select the desired averaging interval.
Press dial knob to confirm.
The selected averaging interval is marked with an arrow.
Press dial knob to return to the main menu.
Press dial knob again to return to the measurement display.
Setting the Volume of the Pulse Tone/Alarm
Tone
The volume for the pulse indicator is selectable from 0 to 9.
The volume for the alarm is selectable from 1 to 4.
Upon switching Vamos on, the default settings are "2" for
Pulse and "1" for Alarm
Turn or press dial knob to select main menu.
Turn dial knob to select »Sound« in the main menu.
Press dial knob to confirm.
In the »Sound« menu, turn dial knob to select item »Pulse«
or »Alarm«.
Press to confirm item.
Turn dial knob to set new value.
Press dial knob to confirm.
Press dial knob to return to the main menu.
Press dial knob again to return to the measurement display.
SpO2
slow
Mode
normal
fast
Fi
Fet
Hal
[%]
N2O
[%]
CO2
[%]
[1/min]
[%]
SpO2
Pulse
98
110
0.6
0.1
30
25
0.0
4.5
exit
Pulse 1
Fi
Fet
Hal
[%]
N2O
[%]
CO2
[%]
[1/min]
[%]
SpO2
Pulse
0.6
0.1
30
25
0.0
4.5
Sound
Alarm
1
98
110
exit
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS

Table of Contents

Related product manuals