EasyManua.ls Logo

Dräger Infinity Delta Series - User Manual

Dräger Infinity Delta Series
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Supplement
Infinity
®
Delta Series
Infinity
®
Patient Monitoring Series
Software VF10.1
WARNING
To properly use this medical device,
read and comply with the instructions
for use and this supplement.

Table of Contents

Questions and Answers

  • R
    Randy MooneySep 4, 2025
    What does 'RRc @ Out Of Range (High)' mean on Dräger Monitor?
    • R
      Roger MedinaSep 4, 2025
      If the Dräger Monitor displays 'RRc @ Out Of Range (High)', it means that the respiration rate has exceeded the upper measuring range. Check ventilation or spontaneous respiration rate.
  • D
    Dennis MccartySep 18, 2025
    Why does the Dräger Infinity Delta Monitor display 'Duplicate Device Connected'?
    • N
      Natalie GibsonSep 18, 2025
      The Dräger Monitor displays 'Duplicate Device Connected' because multiple CO2 devices are connected. Select the desired CO2 device in the etCO2 parameter menu and disconnect the unwanted CO2 device.
  • J
    Jesse PruittSep 16, 2025
    What does 'CO2 Reduced Accuracy' mean on my Dräger Infinity Delta Monitor?
    • J
      joseochoaSep 16, 2025
      The Dräger Monitor displays 'CO2 Reduced Accuracy' because measurement values have reduced accuracy due to warm-up or because the Dräger mainstream CO2 sensor is becoming too warm. Wait until the warm-up phase has ended before relying on the displayed etCO2 values for clinical evaluation. Correct the ambient clinical environment.
  • C
    Charles CardenasSep 14, 2025
    What does 'CO2 H/W Failure' mean on my Dräger Infinity Delta Monitor?
    • M
      Monica NguyenSep 14, 2025
      The Dräger Monitor displays 'CO2 H/W Failure' due to a loss of communication or failure of the entire Dräger mainstream CO2 module. Check cable connections, unplug and re-plug the Dräger mainstream CO2, or power-cycle or undock and re-dock the patient monitor.
  • A
    Angela BrownSep 12, 2025
    What does 'CO2 Disconnected' mean on my Dräger Monitor?
    • C
      Christopher HerreraSep 13, 2025
      The Dräger Monitor displays 'CO2 Disconnected' because the Dräger mainstream CO2 sensor is unplugged and/or the module is disconnected. Check the sensor and module connections.
  • W
    wespinozaSep 12, 2025
    Why does my Dräger Infinity Delta Monitor say 'CO2 Sensor Too Warm'?
    • J
      Jennifer AdamsSep 12, 2025
      The Dräger Monitor displays 'CO2 Sensor Too Warm' because the Dräger mainstream CO2 sensor is too warm as a result of the ambient clinical environment. Correct the ambient clinical environment.
  • K
    Kimberly SchmidtSep 9, 2025
    What causes 'CO2 Window Occluded' on Dräger Infinity Delta Monitor?
    • I
      ihorneSep 9, 2025
      The Dräger Monitor displays 'CO2 Window Occluded' because the window is contaminated or occluded. Ensure the window is clean.
  • E
    Emily JonesSep 7, 2025
    What does 'CO2 Check Cuvette Type' mean on Dräger Monitor?
    • C
      Cindy BriggsSep 7, 2025
      The Dräger Monitor displays 'CO2 Check Cuvette Type' to indicate reduced accuracy. This can happen if a disposable cuvette has been selected on the monitor but a reusable cuvette has been inserted into the breathing apparatus (or vice versa). Ensure the correct cuvette is used and that the cuvette selected on the monitor is correct. Also, the zero may have drifted, so check zero. High inspiratory CO2 concentration may be the reason, so check ventilation settings, CO2 absorber and patient condition.
  • A
    Alexis NicholsSep 5, 2025
    What to do if the Dräger Infinity Delta displays 'inCO2 @ Out Of Range (Low)'?
    • D
      David GregorySep 5, 2025
      If the Dräger Monitor displays 'inCO2 @ Out of Range (Low)', it indicates that the inspiratory gas concentration has fallen below the lower measurement range. Check zero.
  • M
    Mrs. Kristie WatersAug 20, 2025
    What does 'CO2 Please Zero' mean on my Dräger Monitor and how to fix it?
    • J
      Joseph EvansAug 21, 2025
      If the Dräger Monitor displays 'CO2 Please Zero', it means the zero has drifted. Perform a zeroing operation to correct this.

Summary

Infinity Delta Series Software VF10.1

General Safety Warning

Provides critical safety information for device operation.

Mandatory Reporting of Adverse Events

Instructions on reporting serious adverse events with the product.

Definition of Target Groups

Identifies and describes the intended users of the medical device.

Monitor Setup and Configuration

Cybersecurity and Network Stability

Addresses security vulnerabilities and network operational stability.

Unit Manager Menu Settings

Configuration options for alarm control and emergency response.

Core Monitoring Operations and Management

Network Applications and Connectivity

Details on network connections, data exchange, and IT integration.

Alarm Management and Setup

Guides on alarm requirements, limits, history, and latching.

Non-Invasive Blood Pressure (NBP)

Overview, setup, cuff diagram, and status messages for NBP.

Specific Monitoring Parameters

ECG Signal Processing and Leads

Information on ECG signal processing and available lead configurations.

Dräger Mainstream etCO2 Monitoring

Application, setup, connection, display, and troubleshooting for etCO2.

Pulse Contour Cardiac Output (PiCCO) Monitoring

PiCCO parameters, setup, measurement, and accessory details.

Pulse Oximetry (SpO2) Accuracy

Accuracy specifications for Masimo and Nellcor SpO2 sensors.

Maintenance, Reprocessing, and Technical Data

Preventive Maintenance Schedule and Procedures

Details on maintenance intervals and measures for device components.

Reprocessing Procedures and Disinfectants

Guidelines for reprocessing medical devices and approved disinfectants.

Technical Data and EMC Requirements

Specifications for electromagnetic compatibility and immunity.

Troubleshooting Common Issues

Alarm Management Troubleshooting

Resolving issues related to alarms and disconnect messages.

SpO2 and etCO2 Troubleshooting

Guidance for resolving issues with SpO2 and etCO2 monitoring.

Approved Options and Accessories

ECG Lead Sets Information

Details on standard and dual-pin ECG lead sets.

Infinity PiCCO Pod Kit Contents and Cables

List of items included in the Infinity PiCCO pod kit.

Dräger Infinity Delta Series Specifications

General IconGeneral
TouchscreenYes
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
ConnectivityEthernet, USB
Display TypeTFT LCD
Battery TypeLithium-ion
Power Supply100-240 V AC, 50/60 Hz
AlarmsVisual and audible alarms for all parameters
Data StorageInternal memory
CompatibilityDräger

Related product manuals