Supplement – Infinity
®
Delta Series Software VF10.1 35
Supplement for Infinity Delta Series Software
Calculations
The following table lists the parameter labels that have been updated in the Calculations chapter.
Pulse Oximetry (SpO2)
Updated label Former label
Cstat Cs
inO2 iO2
Pmean MAP
Pplat PAUSE
Re Raw e
Ri Raw i
RR RRv
VTe TVe
WARNING
A pulse oximeter should be considered an early warning device. If a trend towards patient
hypoxemia is observed, blood samples should be analyzed by laboratory instruments to
completely understand the condition of the patient.
WARNING
The pulsations from intra-aortic balloon support can elevate the pulse rate. Verify the pulse rate
of the patient against the heart rate.
WARNING
Risk of patient injury
If the SpO
2 sensor is used in the presence of significant concentrations of dyshemoglobins, such
as carboxyhemoglobin or methemoglobin, measurement accuracy may be reduced.
Do not rely on measurement data if the SpO
2 sensor is used under these conditions.
WARNING
If the SpO2 sensor is used in the presence of elevated levels of bilirubin, measurement accuracy
may be reduced.
Do not rely on measurement data if the SpO
2 sensor is used under these conditions.