Supplement – Infinity
®
Delta Series Software VF10.1 23
Supplement for Infinity Delta Series Software
ECG Setup Menu
Menu selection Description Available Settings
The ECG Options submenu
This submenu allows you to configure the following functions:
Pacer Detection Determines the monitor’s ability to
identify pacemaker pulse. Allows user
to enable/disable pacer detection or
choose more advanced Fusion
selection.
CAUTION: Fusion mode pacer
detection is not intended for use with
large-signal, unipolar pacemakers. It is
intended for use only with bipolar
pacemakers.
In basic mode:
Select ON to enable pacemaker
detection.
Select OFF (default) to disable
pacemaker detection.
Pacer Detection
(continued)
Observe the following:
Select Fusion mode only in
situations where it becomes
necessary to suppress repeated
false asystole and/or false low heart
rate alarms.
Before selecting Fusion mode, be
certain that the patient has a bipolar
pacemaker (external or implanted)
and that it is accurately
programmed as appropriate for that
patient.
Do not use Fusion mode if you are
uncertain as to what type of
pacemaker is being used.
In advanced mode:
Select ON to enable pacemaker
detection.
Select OFF (default) to disable
pacemaker detection.
Select Fusion to enable pacer
detection, but minimize pacer tail
rejection to reduce missed detection of
pseudo-fused paced beats, which
results in false asystole alarms.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Delta Series and is the answer not in the manual?

Dräger Infinity Delta Series Specifications

General IconGeneral
TouchscreenYes
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
ConnectivityEthernet, USB
Display TypeTFT LCD
Battery TypeLithium-ion
Power Supply100-240 V AC, 50/60 Hz
AlarmsVisual and audible alarms for all parameters
Data StorageInternal memory