Supplement for Infinity Delta Series Software
84 Supplement – Infinity
®
Delta Series Software VF10.1
To clean the Infinity PiCCO Pod
1 Disconnect the Infinity PiCCO pod from the bedside monitor.
2 Clean the Infinity PiCCO pod with a gauze pad moistened with soapy water or with an approved
cleaning agent.
3 Dry thoroughly with a lint-free cloth.
To disinfect the Infinity PiCCO Pod
1 Disconnect the Infinity PiCCO pod from the bedside monitor.
2 Disinfect the Infinity PiCCO pod with a gauze pad moistened with diluted alcohol.
3 Dry thoroughly with a lint-free cloth.
etCO2
Dräger Mainstream CO2 Module, Sensor, and Cuvettes
Mainstream Module
To clean the Mainstream Module
1 Disconnect the sensor cable (SpO
2 cable, if using) from the Dräger etCO2 mainstream module.
2 Wipe the module surfaces with a cloth moistened with a soap solution.
3 Dry thoroughly with a lint-free cloth.
To disinfect the Mainstream Module
1 Disconnect the sensor cable (including the SpO
2 cable, if using) from the Dräger etCO2 Mainstream
Module.
2 Wipe the module surfaces with a gauze moistened with one of the following:
A 1:3 solution of alcohol
A 1:10 solution of sodium hypochlorite (household bleach)
3 Dry thoroughly with a lint-free cloth.
Mainstream Sensor and Cuvettes
For cleaning and disinfecting instructions for the sensor and reusable cuvettes, refer to the Dräger Infinity
MCable – Mainstream CO
2 IFU.
WARNING
To reduce the risk of infection, remember that the disposable cuvettes are for single patient use
only and cannot be sterilized.