Supplement – Infinity
®
Delta Series Software VF10.1 93
Supplement for Infinity Delta Series Software
System Components
Delta/Delta XL base unit
Kappa CPU base unit
Electrical specifications
Alarm tone sound pressure L(A) alarm
tone sequence
Free field measurements (complying with ISO 3744)
IEC/Infinity
High-priority alarms - 56 dB(A) to 69 dB(A) / 56 dB(A) to
76 dB(A)
Medium-priority alarms - 56 dB(A) to 69 dB(A) / 54 dB(A) to
72 dB(A)
Low-priority alarms - 52 dB(A) to 65 dB(A) / 41 dB(A) to
58 dB(A)
NOTE
The Infinity alarm volume must be set at 20% or higher to
comply with the minimum audio sound pressure of 45 db
specified in standard 60601-2-49.
Environmental Requirements
Temperature range: Operating: 10 to 45 °C (50 to 113 °F)
Storage: –15 to +45 °C (–5 to +113 °F)
NOTE
Storage of the monitor at 40 °C (104 °F) for an extended period
of time (3 - 5 months) degrades the battery capacity.