Quick Reference Table--Apnea Alarm submenu
The following table explains apnea alarm submenu
functions.
Smart Apnea
In OR mode only, the monitor automatically engages the Smart Apnea feature for RRc* apnea alarms if
RRc Apnea Time is set to a selection other than OFF in the etCO2 parameter menu.The Smart Alarm
feature escalates an apnea alarm in the following manner:
To illustrate, if RRc Apnea Time is set to 10 seconds and if an apnea alarm condition is present for
10 seconds, an attention tone annunciates and the alarm message RRc @ Apnea displays. If the apnea
alarm condition continues for 20 seconds, another attention tone annunciates and the alarm message
remains on the monitor. At 30 seconds, a medium-priority alarm annunciates. At 60 seconds a high-
priority alarm annunciates.
Capnograms
The monitor also displays an instantaneous CO2 waveform or capnogram.
Function Description Settings
RRc Apnea Time Specifies time monitor waits before reporting a
cessation of breathing as an apnea event.
OFF, 10, 15, 20, 25, and
30 s
Apnea Archive Allows you to automatically store and/or record an
alarm event for apnea. You can later review store
alarms on the Event Recall screen.
OFF, Record, Store (de-
fault), Str./Rec.
RRc Apnea Time in
Seconds
Alarm Priority Alarm Message Displayed Alarm History
RRc apnea time None; Single attention tone RRc @ Apnea No
RRc apnea time * 2 None; Single attention tone RRc @ Apnea No
RRc apnea time * 3 Medium RRc @ Apnea Yes
RRc apnea time * 6 High RRc @ Apnea Yes
NOTE
If the monitor detects multiple CO2 devices, the first device that was connected to the monitor retains
priority until the user selects a different CO2 device in the etCO2 parameter menu or the undesired
CO
2 is physically disconnected from the monitor.