Supplement for Infinity Delta Series Software
48 Supplement – Infinity
®
Delta Series Software VF10.1
Scale Sets the etCO2* waveform scale. 40, 60, 80, 100 mmHg (40 mmHg
default)
5, 8, 12, 16 kPa (5 kPa default)
5, 8, 12, 16% (5% default)
Resp. Sweep
Speed
Sets the respiratory waveform sweep speed on
the screen’s display.
6.25, 12.5, 25, 50 mm/s
(6.25 mm/s default)
Agent Display Displays a separate agent parameter box.
Ghosted if the monitor is displaying combined
MultiGas parameter box.
–ON
–OFF
Pressure Comp. Sets the compensation for ambient atmospheric
pressure.This item is always set to auto and is
ghosted.
–Auto
RRc* Apnea
Time
Sets the time that the monitors waits before
reporting a cessation of breathing as an apnea
event.
OFF, 10, 15, 20, 25, 30 s
Apnea Archive Stores and/or records an alarm event
automatically for apnea. You can later review
stored alarms on the Event Recall screen.
OFF, 10, 15, 20, 25, 30 s
(Adult: OFF default)
(Pediatric/neonatal: 15 s default)
MultiGas Zero Manually zeros the SCIO module (parameter
values disappear during zeroing).
Not applicable
Auto Zero Delay Delays automatic zeroing for 5 minutes for
uninterrupted monitoring.
WARNING: Delaying the auto zero may
impact accuracy.
NOTE: Gas sensors in the SCIO module are
automatically zeroed and calibrated against
room air. Parameter values disappear during
zeroing. One minute before automatic zeroing,
the monitor sounds an attention tone and
displays the message Auto zero in <1 minute.
Not applicable
etCO2* Alarm Accesses etCO2* alarms in Alarm Limits table (see Chapter 5 of the main IFU)
WARNING
etCO2* alarms do not activate until the first breath is detected after turning on
the monitor or discharging a patient.
Menu Item Description Settings