Supplement – Infinity
®
Delta Series Software VF10.1 19
Supplement for Infinity Delta Series Software
Alarms
General Requirements - Alarm Setup
Upper and Lower Alarm Limits
In addition to those listed in the main IFU, alarm limits should be set according to your patient’s prevailing
condition within the predefined ranges of the monitor as listed in the following table:
Parameter Predefined Alarm Range and
Resolution
Default State Default Alarm Setting
inCO2/inCO2* 2 to 10 mmHg (0.3 to 1.3 kPa) (upper
limit only) increments of 1 mmHg or
0.1 kPa
ON High: 4 mmHg (0.5 kPa)
PLS 30 to 240 in increments of 5 bpm ON Adult:
Low 45 bpm
High 120 bpm
Pediatric:
Low 50 bpm
High 150 bpm
Neonatal:
Low 80 bpm
High 180 bpm