Supplement for Infinity Delta Series Software
20 Supplement – Infinity
®
Delta Series Software VF10.1
Auto Set Alarm Limits
Alarm Limits Table
Function Description Available Settings
Auto Set Sets alarm limits based on current values: Not applicable
NOTE:
The monitor recalculates the upper
and lower alarm limits based on the
parameter values in the Current
column.
Auto Set applies to all displayed
parameters and ST parameters
only.
If a calculated limit value falls
outside the range for that
parameter, the parameters alarm
limits remain unchanged.
Parameters Upper Limit Lower Limit
Ta, T1a, T1b,
T2a, T2b, T3a,
T3b
<
107% of
current value
< 93% of
current value
ΔT1, ΔT2, ΔT3,
PVC/min,
inCO2/inCO2*
No change No change
SpO2/SpO2* Adults 100
Neonates 98
Current value
(value x 5%)
ΔSpO2% Current value
+20%
None
ST Current value
+2.0 mm
Current value
2.0 mm
MultiGas Agent < 105% of
current value
< 95% of
current value
BIS No change No change
TOF-Cnt No change No change
PCCO Current value
+30%
Current value
30%
PCCI Current value
+30%
Current value
30%
MultiGas O2 100% 21%
tp <
125% of
current value
< 80% of
current value
All others Current value
+25%
Current value
20%