Supplement – Infinity
®
Delta Series Software VF10.1 15
Supplement for Infinity Delta Series Software
Adding new devices to the network
Performing upgrades or updates on devices that are connected to the network
Information about connecting to an IT network
Prerequisites
This device must only be connected to the network by service personnel. The IT representative of the
hospital must be consulted in advance.
The following documents must be observed:
Accompanying documents of this device
Descriptions of the network
Description of the network-based alarm systems
Dräger recommends complying with IEC 80001-1 (risk management for IT networks with medical
devices).
LAN networks
LAN networks are usually configured in a star topology. Individual devices can be combined into groups
by means of layer-n-switches. Other data traffic is decoupled by means of separate VLAN networks.
Configure the network settings of the device in accordance with these instructions for use and the network
specifications.
Specifications for LAN connections are described in the following standards:
Wired networks: IEEE˽802.3
Wireless networks: IEEE˽802.11 (b, g)
If the device is used with a layer-2-switch or a layer-3-switch, the port settings must be configured on
the network switch. Before the device is shipped, Dräger can configure the network settings of the
device so that they are compatible with the specifications of the operating organization.
This device exchanges data with other medical devices over the LAN network. The network must
support the following transmissions and protocols:
TCP/IP
Unicast (static or dynamic addressing with the ARP or RARP network protocols)
Multicast
Broadcast
IGMP (version 2)
This device can join or leave an IP multicast group by using the IGMP network protocol.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Delta Series and is the answer not in the manual?

Dräger Infinity Delta Series Specifications

General IconGeneral
TouchscreenYes
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
ConnectivityEthernet, USB
Display TypeTFT LCD
Battery TypeLithium-ion
Power Supply100-240 V AC, 50/60 Hz
AlarmsVisual and audible alarms for all parameters
Data StorageInternal memory
CompatibilityDräger

Related product manuals