Supplement for Infinity Delta Series Software
116 Supplement – Infinity
®
Delta Series Software VF10.1
Pulse Oximetry (SpO2)
Quick Reference Table – SpO2 Setup
If double-pulsing occurs when Pulse Tone Volume is enabled under the SpO2 parameter menu, disable
the pulse tone volume for SpO2 and enable the pulse tone volume in the ECG Options menu.
Dräger Mainstream etCO2 Monitoring
Precautions
Multiple CO2* Devices
In a multiple etCO2* devices situation involving MIB and Scio, it is recommended to physically disconnect
the undesired device from the monitor.
When the message Duplicate Device Connected appears and if the CO2* waveform label disagrees with
the parameters being monitored (e.g., the “CO2* module” waveform label displays when MIB is the active
CO2* source device), disconnect one of the CO2* devices, wait a few second, then reconnect the CO2*
source device.
NOTE
If a MultiGas parameter box appears after a Restore Setup with a Dräger mainstream CO2 module
connected to the monitor, do the following:
1 Disconnect the Dräger mainstream CO2 module.
2 Connect a Scio module.
3 Disconnect the Scio module.
4 Reconnect the Dräger Mainstream CO2 module.
This procedure should remove the MultiGas parameter box.