Supplement – Infinity
®
Delta Series Software VF10.1 37
Supplement for Infinity Delta Series Software
Quick Reference Table - SpO2 Setup
New Status Message
The following status message applies to any monitor using a Masimo SET pod.
MicrO2+ Standalone Pulse Oximeter
As of the Infinity Delta Series Software VF10.1 release, MicrO2+ is no longer supported.
Menu item Description Settings
Pulse Tone
Volume
Sets the volume of the
pulse tone.
NOTE: You can also set
the pulse tone volume
from the ECG menu.
OFF, 5, 10, 20, 30, 40, 50, 60, 70, 80, 90, or 100
SpO2 Alarm Accesses SpO2 alarms
on the Alarm Limits
table.
–SpO2 OxiSure = 20 – 100% in increments of 1%
– Pulse OxiSure = 30 – 240% in increments of 5%
–SpO
2 Masimo = 20 – 100% in increments of 1%
– Pulse Masimo = 30 – 240% in increments of 5%
–SpO
2 Nellcor = 20 – 100% in increments of 1%
– Pulse Nellcor = 30 – 240% in increments of 5%
Message Possible Cause Suggested Action
Replace SpO2 Sensor/Cable The sensor or the cable is
expired or defective.
Replace the sensor or the cable.