Supplement for Infinity Delta Series Software
24 Supplement – Infinity
®
Delta Series Software VF10.1
Arrhythmia Monitoring
Arrhythmia Ranges and Defaults
Parameter Alarm Priority
Default
Rate (default) Count (default) Alarm Archive Factory
Default Setting
ASY HIGH Not adjustable Not adjustable Str./Rec.
VF HIGH Not adjustable Not adjustable Str./Rec.
VT HIGH 100 to 200 (120)
Increments of 10
5 to 15 (10)
Increments of 1
Str./Rec.
RUN MED Not adjustable (VT
rate)
Not adjustable (3 to
VT rate - 1)
Str./Rec.
AIVR MED Not adjustable = (VT
rate - 1)
Not adjustable (3) OFF
SVT MED 120 to 200 (150)
Increments of 10
3 to 10 (3)
Increments of 1
Str./Rec.
TACH MED 100 to 200 (130)
Increments of 10
5 to 15 (8)
Increments of 1
OFF
BRDY MED 30 to 105 (50)
Increment of 5
Not adjustable (8) OFF
CPT LOW Not adjustable Not adjustable Str./Rec.
BGM LOW Not adjustable Not adjustable Str./Rec.
PAUS OFF 1 to 3.5 (2.5)
Increments of 0.5
Not adjustable OFF
ARTF OFF Not adjustable Not adjustable OFF
NOTE
When in neonatal mode or pediatric mode with Basic Arrhythmia monitoring, bradycardia is a high priority
heart rate alarm that is adjustable.