Supplement – Infinity
®
Delta Series Software VF10.1 115
Supplement for Infinity Delta Series Software
Trends
Mini-Trends
When viewing Mini-Trend data and removing and re-connecting parameter cables (e.g., ECG MultiMed
cable), allow the monitor 10 seconds to update Mini-Trend data so that the trend data color matches that
of the monitored parameters.
ECG and Heart Rate
Status Messages
If an ECG Leads Invalid alarm remains displayed even though ECG leads are connected and the alarm
has been acknowledged, disconnect and reconnect the ECG leads to clear the ECG Leads Invalid alarm.
Respiration
OxyCRG (OCRG) Monitoring
Do not disconnect and reconnect the Infinity etCO2 Microstream SmartPod when the OCRG menu is
open, as the trim knob and hard keys may freeze for approximately 30 seconds. After the 30 seconds, the
monitor resumes normal operation.
Cursor
In the OCRG screen, if inaccurate measurement values are suspected when a cursor rests on a point on
a waveform, allow the OCRG screen to populate data for a full 3 minutes if the configured time duration
is 3 minutes (or 6 minutes if the configured time duration is 6 minutes).
Accessing OxyCRG Review Summary Screen
When reviewing the OxyCRG Review Summary screen, if the event recall data appears unavailable,
access the Event Recall screen directly via the Main Menu or Fast Access menus to determine if the
data is truly unavailable.
Time Base
If the OCRG time base is changed (e.g., from 6 minutes to 3 minutes), the OCRG cursor time may be one
minute ahead of the monitor's time. This issue is rectified upon patient discharge.
Recordings
If a timed recording is performed with OCRG so that an etCO2 or etCO2* waveform is printed, the
etCO2/etCO2* waveform appears as a flat line in the middle of the grid. All etCO2-related numeric values
are correct and available.