The suggested action is to disconnect the duplicate device.
Display Features
etCO2 Monitoring
The etCO2 waveform displays the instantaneous CO
2 concentration. The etCO2 parameter box displays
the following parameters:
– End-tidal CO
2 (etCO2) — The level of CO2 in the airway at the end of expiration.
– Inspired CO
2 (inCO2) — The level of CO2 in the airway during the inspiration phase.
– Respiration Rate (RRc) — The patient’s respiration rate, derived from the etCO2 signal by
calculating a n average rate over the two most recent breaths.
etCO2 Param eter Box
Typical etCO2 parameter box displays are shown below.
A
etCO2 (end-tidal CO2) label and value
B
etCO2 alarm limits
C
inCO2 (inspired CO2) label and value
D
inCO2 alarm limit (upper limit only)
E
Respiration rate label and value
F
Respiration rate alarm limits
NOTE
If alarms are disabled, crossed-out triangle icons appear next to the corresponding parameter values.