Supplement for Infinity Delta Series Software
46 Supplement – Infinity
®
Delta Series Software VF10.1
etCO2 Setup
To access the etCO2 setup menu
Click on the etCO2 parameter box.
or
1 Press the Menu fixed key.
2 Click on Patient Setup.
3 Click on Parameters to display a list of available parameters.
4 Click on etCO2.
Quick Reference Table - etCO2 Setup
Click on the etCO2 parameter box to execute
etCO2 setup functions:
Menu Item Description Settings
Source Describes what CO2 device is connected and
displays parameters on the monitor.
NOTE
If one CO2 device is connected to the
monitor, the setting automatically displays
the appropriate device (read-only).
If two CO2 devices are connected, the
settings are made available to the user to
select the desired CO2 device.
(See also the “Multiple CO2 Devices” section.)
–SCIO
Mainstream
Scale Sets the etCO2 waveform scale. 40, 60, 80, 100 mmHg (40 mmHg
default)
5, 8, 12, 16 kPa (5 kPa default)
5, 8, 12, 16% (5% default)
Resp. Sweep
Speed
Sets the respiratory waveform sweep speed on
the screen’s display.
6.25, 12.5, 25, 50 mm/s
(6.25 mm/s default)
Gas
Compensation
Determines gas percentages for inspiratory and
expiratory breathing.
Air (default)
N2O/O2 (in OR mode only)
–O2>60%
HeliOx
Atm. Pressure Sets the compensation for ambient atmospheric
pressure.
540-800 mmHg in increments of
1 mmHg (760 mmHg default)
72.0-106.7 kPa in increments of
0.1 kPa (101.3 kPa default)
Cuvette Type Sets the cuvette type for adult and pediatric
adapters.
Disposable
Reusable (default)