Supplement – Infinity
®
Delta Series Software VF10.1 83
Supplement for Infinity Delta Series Software
Surface disinfection with cleaning
1 Remove soiling immediately. Use a cloth dampened with disinfectant to remove soiling.
2 Perform surface disinfection.
3 After the product has been exposed to the disinfectant for the specified contact time, remove residual
disinfectant.
4 Wipe with a cloth dampened with water (preferably drinking-water quality). Allow the product to dry.
5 Check the product for visible soiling. Repeat steps 1 to 5 if necessary.
6 Check the product for visible damage and replace if necessary.
Cleaning and Disinfecting
IBP - Infinity PiCCO Pod
Agents tested by Dräger and shown to have no harmful effect at the time of testing on the materials utilized
in the Infinity PiCCO Pod include:
Isopropyl alcohol 40%
Higher concentrations could damage:
Compliance™; this cleaning agent may discolor soft plastic material
–Sporox II
Dismozon pur
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
Damage to the device
Electric shock
Device malfunctions
Ensure that no liquid penetrates the device.
CAUTION
Do not autoclave the Infinity PiCCO Pod.
CAUTION
If using alcohol, it should only be a 40% diluted solution.