Supplement – Infinity
®
Delta Series Software VF10.1 17
Supplement for Infinity Delta Series Software
Data have incorrect time stamps.
Requirements for the electrical characteristics of connected devices and networks
The LAN interfaces and the serial interfaces are only suitable for the connection of devices or networks
that have a rated voltage of at most 24 V DC on the network side and that meet the requirements of one
of the following standards:
IEC 60950-1: Ungrounded SELV circuits
IEC 60601-1 (as of 2nd edition): Touchable secondary circuits
Disconnecting the monitor from the network
1 Hold the monitor firmly by its handle. Slide the lever to the left to disengage the power supply. The mon-
itor automatically switches to battery power.
2 Continue to move the lever to the left until it clicks. Use both hands to tilt the monitor forward and lift it
off the IDS or Docking Station.
3 When docking and undocking the monitor to and from an IDS, a minimum 15 second delay between
undocking and redocking the monitor is required, for loading the default IDS profile to the monitor.
Power Sources
Battery Power
The following information refers to the table on battery charge display as described in the main IFU.
NOTE
When the internal battery has approximately 20 minutes remaining, a low priority alarm sounds and
the message Monitor Internal Battery Low appears. The alarm can be silenced; however, the
banner message remains visible.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Delta Series and is the answer not in the manual?

Dräger Infinity Delta Series Specifications

General IconGeneral
TouchscreenYes
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
ConnectivityEthernet, USB
Display TypeTFT LCD
Battery TypeLithium-ion
Power Supply100-240 V AC, 50/60 Hz
AlarmsVisual and audible alarms for all parameters
Data StorageInternal memory
CompatibilityDräger