Supplement for Infinity Delta Series Software
Troubleshooting
The following table describes alarms in alphabetical order by alarm message:
Status Messages
Alarm priority Alarm Cause Remedy
Low 2 Mixed Agents A second anesthetic
agent has been detected.
Wait for the transition phase
to end after changing
anesthetic agents.
Check vaporizer filling level.
Flush system if necessary.
Check fresh gas settings.
Medium 3 Mixed Agents
NOTE
The simultaneous
presence of 3 agents
cannot always be
identified.
A mixture of more than
two anesthetic agents has
been detected.
Wait for the transition phase
to end after changing
anesthetic agents.
Check vaporizer filling level.
Flush system if necessary.
Check fresh gas settings.
Medium Inspired MAC High Inspiratory anesthetic gas
concentration exceeds
3 MAC for a period of 30
or more seconds.
Check vaporizer and fresh
gas settings.
Low MultiGas Module
Disconnected
Gas analyzer is
disconnected.
Check cable connections.
If the problem persists, call
DrägerService.
Medium inO2 > # Inspiratory O2
concentration has
exceeded the upper
alarm limit
Check O2 concentration and
fresh-gas settings
Message Condition Suggested action
Duplicate Device
Connected
Multiple CO
2* devices are
connected.
Select the desired CO2* device in the
etCO2* Parameter menu.
Disconnect the unwanted, duplicate CO
2*
device.