Supplement – Infinity
®
Delta Series Software VF10.1 59
Supplement for Infinity Delta Series Software
Troubleshooting
The following table describes alarms in alphabetical order by alarm message:
Status Messages
Alarm priority Alarm Cause Remedy
Low 2 Mixed Agents A second anesthetic
agent has been detected.
– Wait for the transition phase
to end after changing
anesthetic agents.
– Check vaporizer filling level.
– Flush system if necessary.
– Check fresh gas settings.
Medium 3 Mixed Agents
NOTE
The simultaneous
presence of 3 agents
cannot always be
identified.
A mixture of more than
two anesthetic agents has
been detected.
– Wait for the transition phase
to end after changing
anesthetic agents.
– Check vaporizer filling level.
– Flush system if necessary.
– Check fresh gas settings.
Medium Inspired MAC High Inspiratory anesthetic gas
concentration exceeds
3 MAC for a period of 30
or more seconds.
– Check vaporizer and fresh
gas settings.
Low MultiGas Module
Disconnected
Gas analyzer is
disconnected.
– Check cable connections.
– If the problem persists, call
DrägerService.
Medium inO2 > # Inspiratory O2
concentration has
exceeded the upper
alarm limit
Check O2 concentration and
fresh-gas settings
Message Condition Suggested action
Duplicate Device
Connected
Multiple CO
2* devices are
connected.
– Select the desired CO2* device in the
etCO2* Parameter menu.
– Disconnect the unwanted, duplicate CO
2*
device.