Supplement for Infinity Delta Series Software
54 Supplement – Infinity
®
Delta Series Software VF10.1
Status Messages
Message Condition Suggested action
CO2 Please Zero Zero has drifted Perform a zeroing operation
CO2 Reduced
Accuracy
Measurement values have reduced
accuracy due to warm-up or because the
Dräger mainstream CO
2 sensor is
becoming too warm.
NOTE
Measured parameter labels display
but are followed by a “?”.
Alarm history displays CO2 Reduced
Accuracy. When this condition is
removed, the alarm history displays
CO2 Reduced Accuracy Ended
Parameter box (Pbox) question marks
(?) are not transmitted/displayed over
the network. They only appear in
Pboxes.
Wait until the warm-up phase has
ended before relying on the
displayed etCO2 values for
clinical evaluation.
Correct ambient clinical
environment.
If the problem persists, call
DrägerService.
Duplicate Device
Connected
Multiple CO
2 devices are connected. Select desired CO2 device in the
etCO2 parameter menu.
Disconnect the unwanted CO
2
device.