Supplement for Infinity Delta Series Software
8 Supplement – Infinity
®
Delta Series Software VF10.1
Unit Manager Menu
Menu Item Description Available Settings
The Alarm Control 1 and Alarm Control 2 submenus:
This menu allows the unit manager to configure alarm annunciation. Open the Unit Manager menu,
click on Alarm Control 1 or Alarm Control 2, and then follow the procedures outlined in this table to
execute the indicated functions.
All Alarms OFF
Reminder tone
(Alarm Control 1
submenu)
A reminder tone to indicate all alarms are
OFF permanently.
– Enabled (default)
–Disabled
Enabled when the All Alarms OFF
feature is set to Enabled. A
reminder tone sounds every
30 seconds at 50% volume.
Audio OFF
Reminder (Alarm
Control 1
submenu)
Allows you to set a reminder when the Alarm
Volume is set to OFF.
– Enabled (default) - When the
alarm volume is set to OFF, a
reminder tone sounds every
30 seconds at 50% volume.
For monitors in OR mode: At the
end of an alarm silence or an
alarm off period, if the alarm
condition is still active, the
parameter box flashes and a
reminder tone (high, medium, or
low) sounds every 30 seconds
at 50% volume.
– Disabled - There is no reminder
tone when alarm volume is set
to OFF.
The CODE Setup submenu
This menu allows the unit manager to configure the monitor for quick emergency response. Open the
Unit Manager menu, click on CODE Setup, and then select and execute functions as described in this
table.
Audio OFF Allows you to lower the alarm volume to OFF
when you press the Code fixed key.
NOTE: The central station continues to
sound audio alarms after you press the
Code key.
– Yes - Temporarily sets the alarm
volume and the minimum alarm
volume to OFF when you press
the Code key.
– No (default) - The alarm volume
and the minimum alarm volume
settings do not change when
you press the Code key (i.e.,
retains prior settings).